Calibration Methods for Time-to-Digital Converters

时间数字转换器的校准方法

阅读:1

Abstract

In this paper, two of the most common calibration methods of synchronous TDCs, which are the bin-by-bin calibration and the average-bin-width calibration, are first presented and compared. Then, an innovative new robust calibration method for asynchronous TDCs is proposed and evaluated. Simulation results showed that: (i) For a synchronous TDC, the bin-by-bin calibration, applied to a histogram, does not improve the TDC's differential non-linearity (DNL); nevertheless, it improves its Integral Non-Linearity (INL), whereas the average-bin-width calibration significantly improves both the DNL and the INL. (ii) For an asynchronous TDC, the DNL can be improved up to 10 times by applying the bin-by-bin calibration, whereas the proposed method is almost independent of the non-linearity of the TDC and can improve the DNL up to 100 times. The simulation results were confirmed by experiments carried out using real TDCs implemented on a Cyclone V SoC-FPGA. For an asynchronous TDC, the proposed calibration method is 10 times better than the bin-by-bin method in terms of the DNL improvement.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。