A Highly Sensitive Light-Induced Thermoelastic Spectroscopy Sensor Using a Charge Amplifier to Improve the Signal-to-Noise Ratio

一种利用电荷放大器提高信噪比的高灵敏度光致热弹性光谱传感器

阅读:2

Abstract

A highly sensitive light-induced thermoelastic spectroscopy (LITES) sensor employing a charge amplifier (CA) is reported for the first time in this invited paper. CA has the merits of high input impedance and strong anti-interference ability. The usually used transimpedance amplifier (TA) and voltage amplifier (VA) were also studied under the same conditions for comparison. A standard commercial quartz tuning fork (QTF) with a resonant frequency of approximately 32.76 kHz was used as the photothermal signal transducer. Methane (CH(4)) was used as the target gas in these sensors for performance verification. Compared to the TA-LITES sensor and VA-LITES sensor, the reported CA-LITES sensor shows improvements of 1.83 times and 5.28 times in the minimum detection limit (MDL), respectively. When compared to the LITES sensor without an amplifier (WA-LITES), the MDL has a 19.96-fold improvement. After further optimizing the gain of the CA, the MDL of the CA-LITES sensor was calculated as 2.42 ppm, which further improved the performance of the MDL by 30.3 times compared to the WA-LITES. Additionally, long-term stability is analyzed using Allan deviation analysis. When the average time of the sensor system is increased to 50 s, the MDL of the CA-LITES sensor system can be improved to 0.58 ppm.