All-Fiber LITES Sensor Based on Hollow-Core Anti-Resonant Fiber and Self-Designed Low-Frequency Quartz Tuning Fork

基于空芯反谐振光纤和自主设计的低频石英音叉的全光纤LITES传感器

阅读:1

Abstract

In this paper, an all-fiber light-induced thermoelastic spectroscopy (LITES) sensor based on hollow-core anti-resonant fiber (HC-ARF) and self-designed low-frequency quartz tuning fork (QTF) is reported for the first time. By utilizing HC-ARF as both the transmission medium and gas chamber, the laser tail fiber was spatially coupled with the HC-ARF, and the end of the HC-ARF was directly guided onto the QTF surface, resulting in an all-fiber structure. This design eliminated the need for lens combinations, thereby enhancing system stability and reducing cost and size. Additionally, a self-designed rectangular-tip QTF with a low resonant frequency of 8.69 kHz was employed to improve the sensor's detection performance. Acetylene (C(2)H(2)), with an absorption line at 6534.37 cm(-1) (1.53 μm), was chosen as the target gas. Experimental results clearly demonstrated that the detection performance of the rectangular-tip QTF system was 2.9-fold higher than that of a standard commercial QTF system. Moreover, it exhibited an outstanding linear response to varying C(2)H(2) concentrations, indicating its high sensitivity and reliability in detecting C(2)H(2). The Allan deviation analysis was used to assess the system's stability, and the results indicated that the system exhibits excellent long-term stability.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。