Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes

在健康眼中,与IOLMaster相比,Tomey OA-2000生物测量仪所获得的眼部参数的精确度(重复性和再现性)如何?

阅读:1

Abstract

PURPOSE: To assess the precision (repeatability and reproducibility) of ocular parameters measured by the Tomey OA-2000 biometer, and to compare them with those measured by the IOLMaster. METHODS: In this prospective study, the right eyes of 108 healthy subjects were included. Three consecutive scans were obtained by 2 observers using the Tomey OA-2000, and in the same session one observer used the IOLMaster (version 5.4.4.0006) for the measurements. About 1 week later, 3 scans were obtained by one observer using the Tomey OA-2000. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), keratometer readings, pupil diameter (PD) and corneal diameter (CD) values measured by the Tomey OA-2000 and IOLMaster were analyzed. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within subject standard deviation (Sw) and 2.77Sw were calculated to assess the repeatability and reproducibility. The paired t test and Bland-Altman plots were used to analyze the differences and agreements of parameters measured by the two devices, respectively. RESULTS: Intraobserver repeatability, and interobserver and intersession reproducibility of the AL, CCT, ACD, LT, Kf, Ks, Km, PD and CD values measured by the Tomey OA-2000 biometer showed a CoV of less than 1% except that for PD, and an ICC of more than 0.97 except that for PD and CD. The AL, Kf, Ks, Km and CD values measured by the Tomey OA-2000 were 0.058 ± 0.094 mm, 0.088± 0.150 diopters (D), 0.163 ± 0.170 D, 0.127 ± 0.117 D and 0.171 ± 0.217 mm lower than those measured by the IOLMaster, respectively (all Ps < 0.05). However, the ACD values from the two devices were comparable (P = 0.169). The 95% linite of agreement (LoA) of the AL, ACD, CD and all keratometer readings were no more than 0.24 mm, 0.14 mm 0.60 mm and 0.5 D, respectively. CONCLUSION: Except for the PD and CD, the ocular parameters measured by the Tomey OA-2000 were highly repeatable and reproducible. Except for the CD value, there was good agreement of ocular parameters measured by the Tomey OA-2000 and the IOLMaster in healthy eyes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。