What is the Optimal Nail Length to Treat Osteoporotic Subtrochanteric Fractures? A Finite Element Analysis

治疗骨质疏松性股骨粗隆下骨折的最佳髓内钉长度是多少?有限元分析

阅读:1

Abstract

BACKGROUND: Operative management with intramedullary nail fixation remains the definitive treatment of choice for osteoporotic subtrochanteric (ST) fractures; however, there remains no consensus regarding the proper nail length. We aimed to use 3-dimensional finite element (FE) analysis to determine the optimal nail length for the safe fixation of osteoporotic ST fractures. METHODS: Nine modes of FE models were constructed using 9 different lengths of cephalomedullary nails (short nails: 170, 180, and 200 mm; long nails: 280, 300, 320, 340, 360, and 380 mm) from the same company. The interfragmentary motion was analyzed. Additionally, the peak von Mises stress (PVMS) in the cortical bone, cancellous bone of the femoral head, and the nail were measured, and the yielding risk for each subject was investigated. RESULTS: Long nails were associated with less interfragmentary motion. In the cortical bone, the PVMS of short nails was observed at the distal locking screw holes of the femoral medial cortex; however, in long nails, the PVMS was observed at the lag screw holes on the lateral cortex. The mean yielding risk of long nails was 40.1% lower than that of short nails. For the cancellous bone of the femoral head, the PVMS in all 9 FE models was in the same area: at the apex of the femoral head. There was no difference in the yielding risk between short and long nails. For implants, the PVMS was at the distal locking screw hole of the nail body in the short nails and the nail body at the fracture level in the long nails. The mean yielding risk was 74.9% lower for long nails than that for short nails. CONCLUSIONS: Compared to short nails, long nails with a length of 320 mm or more showed less interfragmentary motion and lower yielding risk in low-level osteoporotic ST fractures. The FE analysis supports long nails as a safer option than short nails, especially for treating transverse-type low-level osteoporotic ST fractures.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。