The FeetMe(®) Insoles System: Repeatability, Standard Error of Measure, and Responsiveness

FeetMe® 鞋垫系统:重复性、测量标准误差和响应性

阅读:2

Abstract

BACKGROUND: Three-dimensional motion analysis using optoelectronic cameras and force platforms is typically used to quantify gait disorders. However, these systems have various limitations, particularly when assessing patients in an ecological environment. To address these limitations, several wearable devices have been developed. However, few studies have reported metrological information regarding their repeatability and sensitivity to change. METHODS: A healthy adult performed 6 min walking tests with FeetMe(®) system insoles under different walking conditions overground and on a treadmill. The standard error of measurement (SEM), the minimum detectable differences (MDDs), and the effect size (ES) were calculated for spatio-temporal parameters, and the ground reaction force was calculated from the 16,000 steps recorded. RESULTS: SEM values were below 3.9% for the ground reaction force and below 6.8% for spatio-temporal parameters. ES values were predominantly high, with 72.9% of cases between overground and treadmill conditions with induced asymmetry, and 64.5% of cases between treadmill conditions with and without induced asymmetry exhibiting an ES greater than 1.2. The minimum detectable differences ranged from 4.5% to 10.7% for ground reaction forces and 2.1% to 18.9% for spatio-temporal parameters. CONCLUSION: Our study demonstrated that the FeetMe(®) system is a reliable solution. The sensitivity to change showed that these instrumented insoles can effectively reflect patient asymmetry and progress.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。