Diagnosis with Metagenomic Next-Generation Sequencing (mNGS) technology and real-time PCR for SARS-CoV-2 Omicron detection using various nasopharyngeal swabs in SARS-CoV-2 Omicron

利用宏基因组二代测序 (mNGS) 技术和实时 PCR 检测 SARS-CoV-2 Omicron,并使用多种鼻咽拭子样本进行诊断

阅读:1

Abstract

BACKGROUND: The SARS-CoV-2 Omicron variant, with the main subtypes BA.5.2 and BF.7 in China, led to off-target effects on the S and N genes from December 1, 2022, to January 31, 2023. The kits used for studying and developing these agents were not adequately and independently evaluated. It is important to verify the performance of commercial Real-Time quantitative PCR (RT-qPCR) tests. OBJECTIVE: We conducted a clinical evaluation of two Real Time SARS-CoV-2 Omicron assays to verify their performance using various detection reagents and clinical specimens. METHODS: We performed clinical evaluations of two existing Chinese SARS-CoV-2 Omicron RT-qPCR kits 2019-nCoV nucleic acid diagnostic kits (Fosun Biotechnology, National instrument registration 20203400299, Shanghai, China) and COVID-19 nucleic acid detection kits (eDiagnosis Biomedicine, National instrument registration 20203400212, Wuhan, China) and using BSD (Bondson) (Guangzhou Bondson Biotechnology Co. Ltd, batch number 2022101), quality controls provided by the inspection center and a large number of clinically confirmed specimens. RESULTS: The concordance rates for the Fosun and eDiagnosis kits were 95% and 100%, respectively. The detection limit for the Fosun and eDiagnosis kits was verified to be 300 copies/mL and 500 copies/mL. The Fosun assay exhibited the largest coefficient of variation (CV) for ORF1ab and N gene at the detection limit concentration (4.80%, 3.49%), whereas eDiagnosis showed a smaller CV (0.93%, 1.10%). In the reference product from the Hangzhou Clinical Laboratory Center test, it was found that Fosun had the lowest sensitivity of 93.47% and a specificity of 100%, while eDiagnosis exhibited 100% for both sensitivity and specificity. The lowest single target gene detection rate of Fosun reagents was 68.7% for the ORF1ab gene and 87.5% for the N gene, while eDiagnosis detection rate was 100%. Among the clinical group S specimens, the missed detection rate of the Fosun reagent was 10.9%, which was higher than the 3.9% of eDiagnosis. However, there was no significant difference in the clinical diagnostic efficiency of the two reagents. CONCLUSIONS: The ORF1ab and N assays of SARS-CoV-2 Omicron on the eDiagnosis platform yielded higher values compared to those on the Fosun platform. Consequently, the eDiagnosis kit has also been used as standard detection reagents. Considering that the Fosun reagent has a relatively low detection limit and targets three single genes, it is more advantageous as a confirmatory reagent for the new museum.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。