ST waveform analysis for monitoring hypoxic distress in fetal sheep after prolonged umbilical cord occlusion

ST波形分析用于监测胎羊长时间脐带阻断后的缺氧窘迫

阅读:1

Abstract

INTRODUCTION: The inconclusive clinical results for ST-waveform analysis (STAN) in detecting fetal hypoxemia may be caused by the signal processing of the STAN-device itself. We assessed the performance of a clinical STAN device in signal processing and in detecting hypoxemia in a fetal sheep model exposed to prolonged umbilical cord occlusion (UCO). METHODS: Eight fetal lambs were exposed to 25 minutes of UCO. ECG recordings were analyzed during a baseline period and during UCO. STAN-event rates and timing of episodic T/QRS rise, baseline T/QRS rise and the occurrence of biphasic ST-waveforms, as well as signal loss, were assessed. RESULTS: During baseline conditions of normoxemia, a median of 40 (IQR, 25-70) STAN-events per minute were detected, compared to 10 (IQR, 2-22) during UCO. During UCO STAN-events were detected in five subjects within 10 minutes and in six subjects after 18 minutes, respectively. Two subjects did not generate any STAN-event during UCO. Biphasic ST event rate was reduced during UCO (median 0, IQR 0-5), compared to baseline (median 32, IQR, 6-55). ST-waveforms could not be assessed in 62% of the recording time during UCO, despite a good quality of the ECG signal. CONCLUSIONS: The STAN device showed limitations in detecting hypoxemia in fetal sheep after prolonged UCO. The STAN device produced high false positive event rates during baseline and did not detect T/QRS changes adequately after prolonged fetal hypoxemia. During 14% of baseline and 62% of the UCO period, the STAN-device could not process the ECG signal, despite its good quality. Resolving these issues may improve the clinical performance of the STAN device.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。