Spinning disk-remote focusing microscopy

转盘远程聚焦显微镜

阅读:11
作者:Michele Gintoli, Sharika Mohanan, Patrick Salter, Elizabeth Williams, James D Beard, Gaspar Jekely, Alexander D Corbett

Abstract

Fast confocal imaging was achieved by combining remote focusing with differential spinning disk optical sectioning to rapidly acquire images of live samples at cellular resolution. Axial and lateral full width half maxima less than 5 µm and 490 nm respectively are demonstrated over 130 µm axial range with a 256 × 128 µm field of view. A water-index calibration slide was used to achieve an alignment that minimises image volume distortion. Application to live biological samples was demonstrated by acquiring image volumes over a 24 µm axial range at 1 volume/s, allowing for the detection of calcium-based neuronal activity in Platynereis dumerilii larvae.