Screening for Major Depressive Disorder Using a Wearable Ultra-Short-Term HRV Monitor and Signal Quality Indices

利用可穿戴式超短期心率变异性监测仪和信号质量指标筛查重度抑郁症

阅读:1

Abstract

To encourage potential major depressive disorder (MDD) patients to attend diagnostic sessions, we developed a novel MDD screening system based on sleep-induced autonomic nervous responses. The proposed method only requires a wristwatch device to be worn for 24 h. We evaluated heart rate variability (HRV) via wrist photoplethysmography (PPG). However, previous studies have indicated that HRV measurements obtained using wearable devices are susceptible to motion artifacts. We propose a novel method to improve screening accuracy by removing unreliable HRV data (identified on the basis of signal quality indices (SQIs) obtained by PPG sensors). The proposed algorithm enables real-time calculation of signal quality indices in the frequency domain (SQI-FD). A clinical study conducted at Maynds Tower Mental Clinic enrolled 40 MDD patients (mean age, 37.5 ± 8.8 years) diagnosed on the basis of the Diagnostic and Statistical Manual of Mental Disorders, Fifth Edition, and 29 healthy volunteers (mean age, 31.9 ± 13.0 years). Acceleration data were used to identify sleep states, and a linear classification model was trained and tested using HRV and pulse rate data. Ten-fold cross-validation showed a sensitivity of 87.3% (80.3% without SQI-FD data) and specificity of 84.0% (73.3% without SQI-FD data). Thus, SQI-FD drastically improved sensitivity and specificity.