Development of a Time Projection Chamber Readout with Hybrid Pixel Sensors for Beam Monitoring

开发用于光束监测的混合像素传感器时间投影室读出装置

阅读:1

Abstract

To monitor the position and profile of therapeutic carbon beams in real-time, in this paper, we proposed a system called HiBeam-T. The HiBeam-T is a time projection chamber (TPC) with forty Topmetal-II- CMOS pixel sensors as its readout. Each Topmetal-II- has 72 × 72 pixels with the size of 83 μm × 83 μm. The detector consists of the charge drift region and the charge collection area. The readout electronics comprise three Readout Control Modules and one Clock Synchronization Module. This Hibeam-T has a sensitive area of 20 × 20 cm and can acquire the center of the incident beams. The test with a continuous 80.55 MeV/u (12)C(6+) beam shows that the measurement resolution to the beam center could reach 6.45 μm for unsaturated beam projections.