Stereo Bi-Telecentric Phase-Measuring Deflectometry

立体双远心相位测量偏转法

阅读:1

Abstract

Replacing the endocentric lenses in traditional Phase-Measuring Deflectometry (PMD) with bi-telecentric lenses can reduce the number of parameters to be optimized during the calibration process, which can effectively increase both measurement precision and efficiency. Consequently, the low distortion characteristics of bi-telecentric PMD contribute to improved measurement accuracy. However, the calibration of the extrinsic parameters of bi-telecentric lenses requires the help of a micro-positioning stage. Using a micro-positioning stage for the calibration of external parameters can result in an excessively cumbersome and time-consuming calibration process. Thus, this study proposes a holistic and flexible calibration solution for which only one flat mirror in three poses is needed. In order to obtain accurate measurement results, the calibration residuals are utilized to construct the inverse distortion map through bicubic Hermite interpolation in order to obtain accurate anchor positioning result. The calibrated stereo bi-telecentric PMD can achieve 3.5 μm (Peak-to-Valley value) accuracy within 100 mm (Width) × 100 mm (Height) × 200 mm (Depth) domain for various surfaces. This allows the obtaining of reliable measurement results without restricting the placement of the surface under test.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。