An Experimental 10-Port Microwave System for Brain Stroke Diagnosis-Potentials and Limitations

用于脑卒中诊断的实验性10端口微波系统——潜力和局限性

阅读:1

Abstract

Microwave imaging systems show potential as replacements for commonly used stroke diagnostic systems. We developed and tested a 10-port microwave system on a liquid head phantom with ischemic and hemorrhagic strokes of varying sizes and positions. This system allows for visualization of changes in dielectric parameters using the TSVD Born approximation, enabling recognition of stroke position and size from the resulting images. The SVM algorithm effectively distinguishes between ischemic and hemorrhagic strokes, achieving 98% accuracy on experimental data, with 99% accuracy in ischemic scenarios and 97% in hemorrhagic scenarios. Using the TSVD Born algorithm, it was possible to precisely image changes in the absolute permittivity of different stroke locations; however, changes in stroke size were more apparent in the variations of absolute permittivity than in the reconstructed stroke size within the antenna plane. Outside this plane, changes in the S-parameters decreased depending on the distance and size of the stroke, making detection and classification more difficult. One ring of antennas around the head proved insufficient, prompting us to focus on developing a system with antennas positioned around the entire head.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。