Hearing Protection Outcomes of Analog Electrode Arrays Coated with Different Drug-Eluting Polymer Films Implanted into Guinea Pig Cochleae

涂有不同药物洗脱聚合物膜的模拟电极阵列植入豚鼠耳蜗的听力保护效果

阅读:5
作者:Yuyu Huang #, Haoran Yu #, Min Liang, Shule Hou, Junmin Chen, Fan Zhang, Xiayu Sun, Huan Jia, Jun Yang

Conclusions

Compared with the negative control, all other groups showed a significant increase in ABR threshold (p<0.001) after surgery. Among them, there was no obvious difference between the blank (0 vs 90 days: 59.70±10.57 vs 64.60±9.47 dB SPL) and the NAD+ group (0 vs 90 days: 59.90±9.87 vs 64.70±8.65 dB SPL). On the other hand, the ABR thresholds in the DXM (0 days: 58.10±10.73 dB SPL; 90 days: 51.70±9.07 dB SPL) and the Ara-C group (0 days: 59.00±10.05 dB SPL; 90 days: 51.60±8.48 dB SPL) decreased significantly compared with the former two groups (p<0.001). However, the Ara-C+DXM group showed no further benefit (p>0.05). In addition, a significantly higher survival rate of spiral ganglion neurons in cochleae was observed in the Ara-C and/or DXM groups.

Methods

Sixty guinea pigs were randomly divided into a negative control group and five experimental groups implanted separately with blank (drug carrier), dexamethasone (DXM), aracytine (Ara-C), Ara-C+DXM, and nicotinamide adenine dinucleotide (NAD+) eluting analog electrode arrays. Micro CT was used to supervise the surgical procedure. Auditory brainstem response (ABR) thresholds of the guinea pigs were measured and analyzed.

Objective

To investigate the hearing protection outcomes of different drug-eluting analog electrode arrays implanted into guinea pig cochleae.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。