Improvement of an Automated Sample Injection System for Pillar Array Columns to Increase Analytical Reproducibility

改进柱阵列色谱柱的自动进样系统以提高分析重现性

阅读:22

Abstract

In our previous study, we developed an automatic sample injection system for pillar array columns for quantitative analysis. An autosampler was used to maintain a constant sample injection volume. However, the sample was diluted during injection using the autosampler, thus deteriorating the analytical reproducibility. In this study, we have substituted the autosampler with a syringe pump to overcome the abovementioned problem and improve the system. Sample dilution was avoided by filling the entire capillary with the sample at a constant rate. This improved system also increased the analytical reproducibility. In the previous system, the relative standard deviation (RSD) exceeded 17% of the peak height for coumarin dyes. In contrast, the improved system decreased the RSD to the range 1.2-1.8%. The analytical reproducibility was evaluated by using five types of amino acids. The RSD of each peak height was within 3.0%, confirming good reproducibility. These results indicate that the sample injection method developed in this study can be applied to biological sample analyses as a simple quantitative analysis method for pillar array columns.