On the extension of the use of a standard operating procedure for nicotine, glycerol and propylene glycol analysis in e-liquids using mass spectrometry

关于扩展使用质谱法分析电子烟液中尼古丁、甘油和丙二醇的标准操作程序

阅读:7
作者:Alessia Turina, Alice Passoni, Silvano Gallus, Alessandra Lugo, Walther Klerx, Reinskje Talhout, Ranti Fayokun, Constantine Vardavas, Enrico Davoli

Conclusions

This study indicates that it is possible to apply the available EU-JATC SOP and the WHO TobLabNet SOP11 even in laboratories that do not have access to an FID, for example, to analyze flavors, trace compounds or carcinogenic, mutagenic, or toxic for reproduction (CMR) in electronic cigarette liquids.

Methods

In May 2020, five test e-liquids in replicates (A-E) were prepared at the Istituto di Ricerche Farmacologiche Mario Negri and sent, with SOP 11, validation document and

Results

The e-liquids analyses, using the two approaches, were compared. Of the 30 laboratories surveyed, 21 utilized the FID approach exclusively, 7 opted for MS detection, and 2 employed both methods. The findings demonstrated that the gas chromatography-mass spectrometry (GC-MS) method offers comparable analytical capabilities regarding accuracy and precision for nicotine, glycerol, and propylene glycol to the FID approach. Through Pearson's correlation test with r≃1 showing a positive correlation between GC-FID and GC-MS data, and the Student's t-test, no significant differences between the two approaches were revealed, showing p>0.005 for almost all three analytes in all samples. Conclusions: This study indicates that it is possible to apply the available EU-JATC SOP and the WHO TobLabNet SOP11 even in laboratories that do not have access to an FID, for example, to analyze flavors, trace compounds or carcinogenic, mutagenic, or toxic for reproduction (CMR) in electronic cigarette liquids.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。