Development and performance evaluation of a rapid in-house ELISA for retrospective serosurveillance of SARS-CoV-2

用于 SARS-CoV-2 回顾性血清监测的快速内部 ELISA 的开发和性能评估

阅读:2
作者:Bijon Kumar Sil, Nowshin Jahan, Md Ahsanul Haq, Mumtarin Jannat Oishee, Tamanna Ali, Shahad Saif Khandker, Eiry Kobatake, Masayasu Mie, Mohib Ullah Khondoker, Mohd Raeed Jamiruddin, Nihad Adnan

Aim

To develop and evaluate a rapid SARS-CoV-2 specific ELISA for detection of anti-SARS-CoV2 IgG from patients' biological samples.

Background

In the ongoing pandemic situation of COVID-19, serological tests can complement the molecular diagnostic

Conclusion

The assay is rapid and can be applied as one of the early and retrospective sero-monitoring tools in all over the affected areas.

Methods

In order to develop this ELISA, three panels of samples (n = 184) have been used: panel 1 (n = 19) and panel 2 (n = 60) were collected from RT-PCR positive patients within 14 and after 14 days of onset of clinical symptoms, respectively; whereas panel 3 consisted of negative samples (n = 105) collected either from healthy donors or pre-pandemic dengue patients. As a capturing agent full-length SARS-CoV2 specific recombinant nucleocapsid was immobilized. Commercial SARS-CoV2 IgG kit based on chemiluminescent assay was used for the selection of samples and optimization of the assay. The threshold cut-off point, inter-assay and intra-assay variations were determined.

Results

The incubation/reaction time was set at a total of 30 minutes with the sensitivity of 84% (95% confidence interval, CI, 60.4%, 96.6%) and 98% (95% CI, 91.1%, 100.0%), for panel 1 and 2, respectively; with overall 94.9% sensitivity (95% CI 87.5%, 98.6%). Moreover, the clinical specificity was 97.1% (95% CI, 91.9%, 99.4%) with no cross reaction with dengue samples. The overall positive and negative predictive values are 96.2% (95% CI 89.2%, 99.2%) and 96.2% (95% CI, 90.6% 99.0%), respectively. In-house ELISA demonstrated 100% positive and negative percent agreement with Elecsys Anti-SARS-CoV-2, with Cohen's kappa value of 1.00 (very strong agreement), while comparing 13 positive and 17 negative confirmed cases.