Small caliber heparin loaded ultrafine microfiber woven graft achieved high patency rate in a preliminary study of canine carotid artery implantation

小口径肝素负载超细纤维编织移植物在犬颈动脉植入的初步研究中取得了较高的通畅率

阅读:10
作者:Masaki Fujita, Nobuaki Tanaka, Yuka Sakaguchi, Yosuke Takaya, Taisuke Kogawa, Hiroshi Tsuchikura, Tadahiro Sasajima, Kazuhiro Tanahashi

Conclusions

The HL-PET graft showed no early postoperative thrombosis and was able to demonstrate a high patency rate with no severe biological response observed after 24 weeks. These results strongly suggest the potential of the HL-PET graft to be used for distal bypasses.

Methods

The HL-PET graft with a diameter of 3 mm and length of 30 mm was made with combination of three key technologies: (1) weaving with PET ultrafine microfiber with a high biological porosity allowing for cell ingrowth, (2) heparin loading on microfiber surfaces, and (3) an outer coating with a flexible bioabsorbable polymer for prevention of blood leakage and graft kinking. Kink resistance, water permeability, and loaded heparin were assessed. One HL-PET graft each was implanted into a carotid artery of six animals. Graft patency rate and healing were assessed 24 weeks after implantation.

Objective

In the past five decades, many small caliber vascular grafts have been developed as bypasses for infrapopliteal or coronary arteries. However, reliable grafts have not been obtained owing to poor patency, mainly caused by early thrombosis or neointimal hyperplasia in the intermediate period after implantation. We developed a novel small caliber heparin-loaded polyethylene terephthalate ultrafine microfiber (HL-PET) graft and evaluated the feasibility to overcome those main causes of graft failure in canine carotid artery implantation.

Results

Among the six grafts, five were deemed patent (patency rate of >83%), with one occluded 20 weeks after implantation. Histopathology of the patent grafts showed neointima formation with confluent endothelial cell lining (estimated mean endothelial cell coverage area, 89 ± 18%). Intimal hyperplasia at the anastomotic sites and severe chronic inflammatory responses were not observed. Immunohistochemistry with antibodies to endothelial nitric oxide synthase, alpha 2 smooth muscle actin and calponin 1 revealed luminal surface endothelial cell layer with expression of endothelial nitric oxide synthase and vascular smooth muscle cells with contractile phenotype in the subintimal layer. Conclusions: The HL-PET graft showed no early postoperative thrombosis and was able to demonstrate a high patency rate with no severe biological response observed after 24 weeks. These results strongly suggest the potential of the HL-PET graft to be used for distal bypasses.