Contemporary total knee designs incorporating highly porous metallic surfaces have demonstrated promising clinical outcomes. However, stiffness differences between modular and monoblock porous tantalum tibial trays may affect bone ingrowth. This study investigated effect of implant design, spatial location and clinical factors on bone ingrowth. Three modular and twenty-one monoblock retrieved porous tantalum tibial trays were evaluated for bone ingrowth. Nonparametric statistical tests were used to investigate differences in bone ingrowth by implant design, tray spatial location, substrate depth and clinical factors. Modular trays (5.3 ± 3.2%) exhibited higher bone ingrowth than monoblock trays (1.6 ± 1.9%, P = 0.032). Bone ingrowth in both designs was highest in the initial 500 μm from the surface. Implantation time was positively correlated with bone ingrowth for monoblock trays.
Is There A Difference in Bone Ingrowth in Modular Versus Monoblock Porous Tantalum Tibial Trays?
模块化多孔钽胫骨托盘与整体式多孔钽胫骨托盘在骨长入方面是否存在差异?
阅读:7
作者:Hanzlik Josa A, Day Judd S, Rimnac Clare M, Kurtz Steven M
| 期刊: | Journal of Arthroplasty | 影响因子: | 3.800 |
| 时间: | 2015 | 起止号: | 2015 Jun;30(6):1073-8 |
| doi: | 10.1016/j.arth.2015.01.010 | 研究方向: | 骨科研究 |
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。