Performance evaluation of the nanoScan(®) P123S total-body PET

nanoScan® P123S 全身 PET 的性能评估

阅读:1

Abstract

PURPOSE: Before utilising preclinical Position Emission Tomography (PET) systems for biological studies, evaluating their performance is important to better qualify the scanner's applications. This study aims to assess the performance of the new extended field of view (FOV) nanoScan® PET/CT P123S system, developed for rodent total-body PET applications. METHODS: Scanner resolution, noise equivalent count rate (NECR), sensitivity and image quality were evaluated following NEMA NU-4 2008 protocols. Furthermore, a Derenzo phantom and linearity measurements were conducted. In vivo studies were subsequently carried out to evaluate system performance in biological applications. RESULTS: The scanner spatial resolution according to the NEMA protocol was 1.4 mm using FBP reconstruction, while with iterative reconstruction it was under 0.7 mm. The NECR peak using a 250‒750 keV energy window was 1805.0 kcps at 93.7 MBq and 880.7 kcps at 88.4 MBq for the mouse-sized and rat-sized phantom respectively. The absolute sensitivity was 10.5%. The standard deviation of the uniform area of the image quality phantom was 1.8%, while the recovery coefficients varied between 0.23 and 1.00. The spill-over ratios were 0.04, and 0.04 in the water and air-filled chambers respectively. Quantitative bias was < 4% with a linear response up to 105 MBq. Total-body rat images were successfully acquired using the new system. CONCLUSION: The new extended FOV PET system has improved sensitivity and count rate performance compared with previous systems. Its spatial resolution and quantitative accuracy are well-suited for preclinical PET applications. The extended FOV enables total-body imaging of both mice and rats.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。