Evaluation of the Potential Risk of Analyzing Phosphopeptides with Easy-nLC 1200-Coupled Mass Spectrometers

使用 Easy-nLC 1200 联用质谱仪分析磷酸肽的潜在风险评估

阅读:5
作者:Xiaoxiao Duan, Yuanya Zhang, Xiahe Huang, Xiao Ma, Hui Gao, Yingchun Wang

Abstract

Liquid chromatography-mass spectrometry (LC-MS) is a major tool for the large-scale qualitative and/or quantitative analysis of protein phosphorylation in cells or tissues. The performance of LC is pivotal for the success of phosphoproteomics in both sensitivity and reproducibility. Here, we report that the widely used Easy-nLC 1200 has poor performance in analyzing phosphopeptides, particularly in terms of sensitivity and reproducibility, whereas its predecessor, Easy-nLC 1000, has a much better performance. Therefore, we suggest that Easy-nLC 1200 is not appropriate for LC-MS-based proteomics analysis for samples with a limited amount, particularly phosphopeptides from plants.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。