SERS-based immunoassay using gold-patterned array chips for rapid and sensitive detection of dual cardiac biomarkers

使用金图案阵列芯片的 SERS 免疫测定法可快速灵敏地检测双重心脏生物标志物

阅读:5
作者:Ziyi Cheng, Rui Wang, Yanlong Xing, Linlu Zhao, Jaebum Choo, Fabiao Yu

Abstract

Cardiac troponin I (cTnI) and creatine kinase-MB (CK-MB) are important diagnostic biomarkers for acute myocardial infarction (AMI). Many efforts have been undertaken to develop highly sensitive detection methods for the quantitative analysis of these dual targets. However, current immunoassay methods are inadequate for accurate measurement of cTnI and CK-MB, due to their limited detection sensitivity. Thus, there is still an urgent demand for a new technique that will enable ultrahigh sensitive detection of these biomarkers. In this study, we developed a surface-enhanced Raman scattering (SERS)-based sandwich immunoassay platform for the ultrasensitive detection of cTnI and CK-MB. In this study, a monoclonal-antibody-immobilized gold-patterned chip was used as a SERS active template. Target samples and polyclonal-antibody-conjugated Au@Ag core-shell nanoparticles were then added. Using this SERS platform, the concentration of biomarkers could be quantified by monitoring the characteristic Raman peak intensity of Raman reporter molecules. Under optimized conditions, the limits of detection (LODs) were estimated to be 8.9 pg mL-1 and 9.7 pg mL-1 for cTnI and CK-MB, respectively. Thus, the proposed SERS-based immunoassay has great potential to be an effective diagnostic tool for the rapid and accurate detection of cTnI and CK-MB.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。