Assessment of retinal vein pulsation through video-ophthalmoscopy and simultaneous biosignals acquisition

通过视频眼底镜检查和同步生物信号采集评估视网膜静脉搏动

阅读:1

Abstract

The phenomenon of retinal vein pulsation is still not a deeply understood topic in retinal hemodynamics. In this paper, we present a novel hardware solution for recording retinal video sequences and physiological signals using synchronized acquisition, we apply the photoplethysmographic principle for the semi-automatic processing of retinal video sequences and we analyse the timing of the vein collapse within the cardiac cycle using of an electrocardiographic signal (ECG). We measured the left eyes of healthy subjects and determined the phases of vein collapse within the cardiac cycle using a principle of photoplethysmography and a semi-automatic image processing approach. We found that the time to vein collapse (T(vc)) is between 60 ms and 220 ms after the R-wave of the ECG signal, which corresponds to 6% to 28% of the cardiac cycle. We found no correlation between T(vc) and the duration of the cardiac cycle and only a weak correlation between T(vc) and age (0.37, p = 0.20), and T(vc) and systolic blood pressure (-0.33, p = 0.25). The Tvc values are comparable to those of previously published papers and can contribute to the studies that analyze vein pulsations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。