Superior reproducibility and repeatability in automated quantitative pupillometry compared to standard manual assessment, and quantitative pupillary response parameters present high reliability in critically ill cardiac patients

与标准手动评估相比,自动定量瞳孔测量法具有更优异的重现性和重复性,定量瞳孔反应参数在危重心脏病患者中表现出较高的可靠性。

阅读:2

Abstract

BACKGROUND: Quantitative pupillometry is part of multimodal neuroprognostication of comatose patients after out-of-hospital cardiac arrest (OHCA). However, the reproducibility, repeatability, and reliability of quantitative pupillometry in this setting have not been investigated. METHODS: In a prospective blinded validation study, we compared manual and quantitative measurements of pupil size. Observer and device variability for all available parameters are expressed as mean difference (bias), limits of agreement (LoA), and reliability expressed as intraclass correlation coefficients (ICC) with a 95% confidence interval. RESULTS: Fifty-six unique quadrupled sets of measurement derived from 14 sedated and comatose patients (mean age 70±12 years) were included. For manually measured pupil size, inter-observer bias was -0.14±0.44 mm, LoA of -1.00 to 0.71 mm, and ICC at 0.92 (0.86-0.95). For quantitative pupillometry, we found bias at 0.03±0.17 mm, LoA of -0.31 to 0.36 mm and ICCs at 0.99. Quantitative pupillometry also yielded lower bias and LoA and higher ICC for intra-observer and inter-device measurements. Correlation between manual and automated pupillometry was better in larger pupils, and quantitative pupillometry had less variability and higher ICC, when assessing small pupils. Further, observers failed to detect 26% of the quantitatively estimated abnormal reactivity with manual assessment. We found ICC >0.91 for all quantitative pupillary response parameters (except for latency with ICC 0.81-0.91). CONCLUSION: Automated quantitative pupillometry has excellent reliability and twice the reproducibility and repeatability than manual pupillometry. This study further presents novel estimates of variability for all quantitative pupillary response parameters with excellent reliability.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。