Dose rate determination for preclinical total body irradiation

临床前全身照射剂量率测定

阅读:1

Abstract

The accuracy of delivered radiation dose and the reproducibility of employed radiotherapy methods are key factors for preclinical radiobiology applications and research studies. In this work, ionization chamber (IC) measurements and Monte Carlo (MC) simulations were used to accurately determine the dose rate for total body irradiation (TBI), a classic radiobiologic and immunologic experimental method. Several phantom configurations, including large solid water slab, small water box and rodentomorphic mouse and rat phantoms were simulated and measured for TBI setup utilizing a preclinical irradiator XRad320. The irradiator calibration and the phantom measurements were performed using an ADCL calibrated IC N31010 following the AAPM TG-61 protocol. The MC simulations were carried out using Geant4/GATE to compute absorbed dose distributions for all phantom configurations. All simulated and measured geometries had favorable agreement. On average, the relative dose rate difference was 2.3%. However, the study indicated large dose rate deviations, if calibration conditions are assumed for a given experimental setup as commonly done for a quick determination of irradiation times utilizing lookup tables and hand calculations. In a TBI setting, the reference calibration geometry at an extended source-to-surface distance and a large reference field size is likely to overestimate true photon scatter. Consequently, the measured and hand calculated dose rates, for TBI geometries in this study, had large discrepancies: 16% for a large solid water slab, 27% for a small water box, and 31%, 36%, and 30% for mouse phantom, rat phantom, and mouse phantom in a pie cage, respectively. Small changes in TBI experimental setup could result in large dose rate variations. MC simulations and the corresponding measurements specific to a designed experimental setup are vital for accurate preclinical dosimetry and reproducibility of radiobiological findings. This study supports the well-recognized need for physics consultation for all radiobiological investigations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。