A Systematic Investigation of Parameters Influencing Droplet Rain in the Listeria monocytogenes prfA Assay - Reduction of Ambiguous Results in ddPCR

对影响单核细胞增生李斯特菌prfA检测中液滴雨的参数进行系统研究——减少ddPCR中的歧义结果

阅读:1

Abstract

The droplet digital polymerase chain reaction (ddPCR) determines DNA amounts based upon the pattern of positive and negative droplets, according to Poisson distribution, without the use of external standards. However, division into positive and negative droplets is often not clear because a part of the droplets has intermediate fluorescence values, appearing as "rain" in the plot. Despite the droplet rain, absolute quantification with ddPCR is possible, as shown previously for the prfA assay in quantifying Listeria monocytogenes. Nevertheless, reducing the rain, and thus ambiguous results, promotes the accuracy and credibility of ddPCR. In this study, we extensively investigated chemical and physical parameters for optimizing the prfA assay for ddPCR. While differences in the concentration of all chemicals and the dye, quencher and supplier of the probe did not alter the droplet pattern, changes in the PCR cycling program, such as prolonged times and increased cycle numbers, improved the assay.