Prospective evaluation of accuracy of HIV viral load monitoring using the Aptima HIV Quant Dx assay with fingerstick and venous dried blood spots prepared under field conditions in Kenya

使用 Aptima HIV Quant Dx 检测法,在肯尼亚现场条件下制备指尖采血和静脉干血斑,对 HIV 病毒载量监测的准确性进行前瞻性评估

阅读:4
作者:Matilu Mwau, Jeff Danda, Joseph Mbugua, Allan Handa, Jacqueline Fortunko, Andrew Worlock, Sangeetha Vijaysri Nair

Abstract

Quantification of HIV-1 RNA is essential for clinical management of HIV patients. The limited throughput and significant hands-on time required by most HIV Viral load (VL) tests makes it challenging for laboratories with high test volume, to turn around patient results quickly. The Hologic Aptima HIV-1 Quant Dx Assay (Aptima), has the potential to alleviate this burden as it is high throughput and fully automated. This assay is validated for both plasma and dried blood spots (DBS), which are commonly used in resource limited settings. The objective of this study was to compare the performance of Aptima to Abbott RealTime HIV-1 Assay (Abbott RT), which was used as reference. This was a cross-sectional prospective study where HIV VL in finger stick (FS) DBS, venous blood (VB) DBS and plasma, collected from 258 consenting adults visiting 5 medical facilities in Kenya, Africa were tested in Aptima. The results were compared to plasma VL in Abbott RT at the medical decision point (MDP) of 1000 copies/mL and across Aptima assay range. The total agreement at MDP between plasma HIV VL in Abbott RT and plasma, FS and VB DBS tested in Aptima were 97.7%, 92.2% and 95.3% respectively with kappa statistic of 0.95, 0.84 and 0.90. The positive and negative agreement for all 3 sample types were >92%. Regression analysis between VL in Abbott RT plasma and various sample types tested in Aptima had a Pearson's correlation coefficient ≥0.91 with systematic bias of < 0.20 log copies/mL on Bland-Altman analysis. The high level of agreement in Aptima HIV VL results for all 3 sample types with Abbott RT plasma VL along with the high throughput, complete automation, and ease of use of the Panther platform makes Aptima a good option for HIV VL monitoring for busy laboratories with high volume of testing.

特别声明

1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。

2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。

3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。

4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。