Task-based automatic keV selection: leveraging routine virtual monoenergetic imaging for dose reduction on clinical photon-counting detector CT()

基于任务的自动keV选择:利用常规虚拟单能成像降低临床光子计数探测器CT的剂量

阅读:2

Abstract

Objective. Photon-counting detector (PCD) CT enables routine virtual-monoenergetic image (VMI) reconstruction. We evaluated the performance of an automatic VMI energy level (keV) selection tool on a clinical PCD-CT system in comparison to an automatic tube potential (kV) selection tool from an energy-integrating-detector (EID) CT system from the same manufacturer.Approach.Four torso-shaped phantoms (20-50 cm width) containing iodine (2, 5, and 10 mg cc(-1)) and calcium (100 mg cc(-1)) were scanned on PCD-CT and EID-CT. Dose optimization techniques, task-based VMI energy level and tube-potential selection on PCD-CT (CARE keV) and task-based tube potential selection on EID-CT (CARE kV), were enabled. CT numbers, image noise, and dose-normalized contrast-to-noise ratio (CNR(d)) were compared.Main results. PCD-CT produced task-specific VMIs at 70, 65, 60, and 55 keV for non-contrast, bone, soft tissue with contrast, and vascular settings, respectively. A 120 kV tube potential was automatically selected on PCD-CT for all scans. In comparison, EID-CT used x-ray tube potentials from 80 to 150 kV based on imaging task and phantom size. PCD-CT achieved consistent dose reduction at 9%, 21% and 39% for bone, soft tissue with contrast, and vascular tasks relative to the non-contrast task, independent of phantom size. On EID-CT, dose reduction factor for contrast tasks relative to the non-contrast task ranged from a 65% decrease (vascular task, 70 kV, 20 cm phantom) to a 21% increase (soft tissue with contrast task, 150 kV, 50 cm phantom) due to size-specific tube potential adaptation. PCD-CT CNR(d)was equivalent to or higher than those of EID-CT for all tasks and phantom sizes, except for the vascular task with 20 cm phantom, where 70 kV EID-CT CNR(d)outperformed 55 keV PCD-CT images.Significance. PCD-CT produced more consistent CT numbers compared to EID-CT due to standardized VMI output, which greatly benefits standardization efforts and facilitates radiation dose reduction.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。