Development and Validation of a Digitizer-Based TCSPC System for Scintillation Decay Time Analysis via an Extended Convolution Model

基于数字化仪的TCSPC系统在闪烁衰减时间分析中的开发与验证——基于扩展卷积模型

阅读:1

Abstract

The development of high-fidelity digital twins for scintillation spectrometer detectors demands precise experimental characterization of timing parameters. This work presents a comprehensive solution comprising a digitizer-based time-correlated single-photon counting (TCSPC) system and an extended convolution model for decay time analysis. We introduce a physics-driven calibration principle, validating the system response against an independent physical benchmark to ensure fidelity. The proposed convolution model advances beyond the conventional model by incorporating additional parameters to account for scintillator-induced timing broadening and delay, thereby decoupling this effect from instrumental response. The model's descriptive power was statistically validated through its application to fast scintillators, while its physical accuracy was robustly confirmed through the precise extraction of typical decay times from slow scintillators. This methodology establishes a reliable workflow from measurement to parameterization, directly supplying the decoupled inputs required for the digital twins of scintillation detectors.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。