Evaluation of type-specific real-time PCR assays using the LightCycler and J.B.A.I.D.S. for detection of adenoviruses in species HAdV-C

使用 LightCycler 和 JBAIDS 进行类型特异性实时 PCR 检测以评估 HAdV-C 种腺病毒

阅读:3
作者:Morris S Jones, Nolan Ryan Hudson, Carl Gibbins, Stephen L Fischer

Abstract

Sporadically, HAdVs from species HAdV-C are detected in acute respiratory disease outbreaks. To rapidly type these viruses, we designed real-time PCR assays that detect and discriminate between adenovirus types HAdV-C1, -C2, -C5, and -C6. Sixteen clinical isolates from the California Department of Public Health were used to validate the new assays. Type-specific TaqMan real-time PCR assays were designed and used independently to successfully identify 16 representative specimens. The lower limit of detection for our LightCycler singleplex real-time PCR assays were calculated to be 100, 100, 100, and 50 genomic copies per reaction for HAdV-C1, HAdV-C2, HAdV-C5 and HAdV-C6, respectively. The results for the singleplex J.B.A.I.D.S. assays were similar. Our assays did not cross-react with other adenoviruses outside of species HAdV-C, respiratory syncytial virus, influenza, or respiratory disease causing bacteria. These assays have the potential to be useful as diagnostic tools for species HAdV-C infection.